Skip Ribbon Commands
Skip to main content

Clinical evaluation of the novel Capiox NX19 adult oxygenator–a multicenter study

5/07/2022 -  

Introduction

The novel Capiox NX19 adult oxygenator is, compared to its predecessors, improved with enhanced air removal technology, a polymer heat exchanger and smaller, innovative hollow fibers resulting in a surface area reduction and a lower priming volume. The aim of this study was to evaluate the NX19 oxygenator performance in a clinical setting.


Methods

A prospective multicenter study was performed involving three large European university hospitals. The Capiox NX19 (n = 150) performance was assessed during adult cardiopulmonary bypass and involved gaseous microemboli handling and gas transfer efficiency. The heat exchanger performance was evaluated separately in vitro.


Results

The heat exchanger performance factors were 0.80 ± 0.03 and 0.58 ± 0.04 at pump flow rates of 3 L/min and 6 L/min, respectively. After priming, residual post-oxygenator gaseous microemboli count and volume were decreased by 91% and 93.7%, respectively. The gas compartment pressure was 6.0 ± 2.5 mmHg, while the O2 transfer was 69 ± 30 mL/min/m2 and the CO2 transfer 73 ± 34 mL/min/m2. The O2 gradient was 44 ± 19 mmHg/LPM and the O2 diffusing capacity 0.38 ± 0.14 mL/min/mmHg. The shunt fraction was 0.19 ± 0.13, whereas oxygenator resistance and shear stress were 10.5 ± 3.7 mmHg/LPM and 5.1 ± 3.1 dyn/cm2, respectively.


Conclusion

This multicenter study displayed good clinical safety and performance of the NX19 oxygenator.


Key findings

  • Capiox NX19 showed improved heat exchanger performance (0,8 @ 3L/min) compared to FX25 (0,74 @ 3L/min)
  • Similar GME removal between Capiox NX19 and Capiox FX25 (93,7 vol% removal) – an excellent result taking into account the smaller size of NX19
  • Higher O2 transfer of NX19 compared to FX25 (69 ± 30 ml/min/m2 vs. 44 ± 14 ml/min/m2)
  • Higher CO2 removal of NX19 compared to FX25 (73 ± 34 ml/min/m2 vs. 26 ± 14 ml/min/m2)
  • Lower pressure drop of NX19 (ΔP = 10,5 mmHg/LPM) compared to most other contemporary oxygenators (Fusion = 16 mmHg/LPM, Inspire 8 = 27 mmHg/LPM, Inspire 6 = 30mmHg/LPM)


Keywords

perfusion, oxygenator, Terumo Capiox NX19, clinical performance, oxygenator evaluation, cardiopulmonary bypass, microemboli, gas transfer, oxygen transfer, carbon dioxide transfer, heat exchanger performance


The opinions expressed by Hendrix Rik H J et al. are their own. Unattributed data, device selection and procedural guidance is a matter of physician preference and opinion derived from the clinicians’ own observations and experience.